Trilogy 100

Ventilator

Versatile, easy-to-use Trilogy 100 is designed for home, hospital and acute- care use to provide invasive and non-invasive ventilator support for adult and pediatric patients. A pressure and volume mode ventilator with proprietary algorithms, Trilogy 100 meets the changing needs of patients by combining unique flexibility with ease of use and advanced technology.

Specifications Visa mer

Environmental - IntelliBridge EC40/80
Storage Temperature
-20°C to 60 °C (-4 °F to 140°F)
Operating temperature
5°C to 40 °C (41 °F to 104°F)
Relative humidity
15 - 95 % (non-condensing)
Atmospheric pressure
110 to 60 kPa
Dimensions and weight
Dimensions
16.68 cm L x 28.45 cm W x 23.52 cm H (6.6" L x 11.2" W x 9.3" H)
Weight
5 kg (11 lb)
Controls
Circuit types
Active with PAP; Passive
Ventilation types
Pressure and Volume
Volume modes
AC; SIMV (w/PS); CV; AC-MPV
Pressure modes
CPAP; S; S/T; T; PC; PC-SIMV (w/PS); PC-MPV; AVAPS-AE
Pressure
4 to 50 cm H₂O
IPAP
4 to 50 cm H₂O
PEEP/EPAP for active circuits
0 to 25 cm H₂O
PEEP/EPAP for passive circuits
4 to 25 cm H₂O
CPAP for passive circuits
4 to 20 cm H₂O
Pressure support differential
0 to 40 cm H₂O
Alarms
Apnea rate
Off; 10 to 60 sec
Circuit disconnect
Off; 5 to 60 sec
High tidal volume
Off; 50 to 2000 ml
Low tidal volume
Off; 40 to 2000 ml
High minute ventilation
Off; 1.0 to 99 l/min
Low minute ventilations
Off; 0.2 to 99 l/min
High respiratory rate
Off; 4 to 80 BPM
Low respiratory rate
Off; 4 to 80 BPM
Measured patient parameters
Respiratory rate
0 to 80 BPM
Minute ventilation
0 to 99 l/min
Vte/Vti
0 to 2000 ml
Leak rate
0 to 200 l/min
Peak inspiratory flow
0 to 200 l/min
Peak inspiratory pressure
0 to 99 cm H₂O
Mean airway pressure
0 to 99 cm H₂O
% Patient triggered breaths
0 to 100 %
I:E ratio
9.9:1 to 1:9.9
Controls
Tidal volume
50 to 2000 ml
Breath rate
0 to 60 BPM for AC mode; 1 to 60 BPM for all other modes
Inspiratory time
0.3 to 5.0 sec
Rise time
1 to 6
Ramp start pressure
0 to 25 cm H₂O for active circuits; 4 to 25 cm H₂O for passive circuits; 4 to 19 cm H₂O in CPAP
Ramp length
Off; 5 to 45 min
Flex
Off; 1 to 3
Flow trigger sensitivity
1 to 9 l/min
Flow cycle
10 to 90 %
Apnea rate
4 to 60 BPM